Performance Verification
23
Measure the peak-to-peak amplitude on the oscilloscope. It should be greater
than or equal to 210 mV – scope vertical accuracy. Record the test results as
Bandwidth in
24
Disable the 9500B's output and disconnect the probe input.
Performance Test Record
Serial Number:
Certification Date:
Certification Temperature:
Tested By:
Table 10
N2804A Performance Test Record
Test
DC Differential Gain
Accuracy
Bandwidth
34
Table 10
Test Limits
±1% (99 mV – scope vertical accuracy to 101 mV + scope vertical accuracy)
¡
(210 mV – scope vertical accuracy)
on page 34.
Test Results
N2804/5A Differential Probes User's Guide