Power Supply Assembly; Removal - Keysight N8973B Service Manual

Nfa series
Hide thumbs Also See for N8973B:
Table of Contents

Advertisement

Assembly Replacement Procedures

Power Supply Assembly

Power Supply Assembly

Removal

1. Remove the instrument outer case. Refer to the
2. Remove the rear panel. Refer to the
3. Remove the top brace and reference bracket. Refer to the
4. Remove the reference assembly. Refer to the
5. Remove the CPU assembly. Refer to the
6. Refer to
7. The power supply assembly can be removed from the chassis by pulling
Figure 15-32
Power Supply Assembly Removal
Keysight NFA Series Noise Figure Analyzers Service Guide
removal procedure.
removal procedure.
removal procedure.
procedure.
Figure
15-32. Remove the three screws (1) attaching the power
supply (2) to the power supply bracket (3).
straight out the back.
"Instrument Outer Case"
"Rear Panel"
removal procedure.
"Reference Assembly"
"CPU Assembly"
removal
"Top Brace"
431

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

N8974bN8975bN8976b