Bubble Sensor Input and Bubble Sensor Output

Bubble Sensor are present on lines B, C, D, E & F to detect whether fluid or air is present in that
section of the tube.
A Bubble Sensor Input trigger detects change using the bubble sensor on the source bag of the fluid
path, while a Bubble Sensor Output trigger detects change using the bubble sensor on the destination
bag of the fluid path.
Use the dropdown menu to select whether you want the trigger condition to be met when the sensor
first detects air (Wet to Dry), or when it first detects fluid (Dry to Wet).
Select the Pause on Bubble check box to put the instrument into a paused state when the trigger
occurs rather than advancing to the next step, requiring a manual use of the skip button to advance.
CTS
Rotea
Counterflow Centrifugation System Process Design User Guide
Chapter 3 Rotea Protocol Builder Application
3
Trigger
37