Timer (Seconds) And Timer (Minutes); Skip Button Triggered - Thermo Scientific gibco CTS Rotea User Manual

Counterflow centrifugation system process design
Table of Contents

Advertisement

Timer (Seconds) and Timer (Minutes)

A Timer (Seconds) or Timer (Minutes) condition causes the loop to continue until the entered length of
time has passed.
Note: The condition is checked at the start of each loop. If the timer value is set to 30 seconds, and it
reaches that time in the middle of the loop, it will finish the current loop before continuing out of it.

Skip Button Triggered

A Skip Button Triggered condition will cause the loop to continue until the Advance button on the
instrument is manually pressed by the user.