The Rotea
software uses triggers to automate the progression from one step to the next in a protocol.
Each step must have at least one valid trigger. A trigger is a condition that, when met, causes the
protocol to advance to the next step.
To create triggers:
1. Select a Trigger type from the dropdown menu.
Note: The Skip Button box at the bottom of the Step view is also a valid trigger as it enables the
user to manually advance to the next step.
Each trigger type has different parameters. These are addressed in the following subsections.
2. Add triggers by clicking
3. If you want the instrument to go into a pause state instead of automatically progressing to the
next step, click Pause on Trigger. Pause on Trigger is only available on Pressure 1, Pressure
2, Bubble Sensor Input and Bubble Sensor Output step types. When this option is used the
instrument will require a manual advance using the Skip Button before the next step will begin.
32
or delete by clicking the
CTS
Rotea
Counterflow Centrifugation System Process Design User Guide
adjacent to the trigger.