Chapter 2 Equipment Requirements

1 System Components

1.1 Presentation of the 3 Rooms

1.1.1 Examination room

Frontal Positioner
Lateral Positioner
Omega Patient Table
Performix X–Ray Tube Assembly (x 2)
Collimator (x 2)
Innova Digital Detector (x 2)

1.1.2 Technical room

C2 Cabinet (Frontal/Lateral)
C1 Frontal Cabinet and C1 Lateral Cabinet
Coolix 4100 chiller (2 Cabinets)
Detector Conditioner HEAT DRY 1 (2 Cabinets)
Fluoro UPS (option) (1 Cabinet)
Power Distribution Box (Main Disconnet Panel)
3 kVA cabinets UPS
Large Display Cabinet and UPS (for Large Display Option)
Chapter 2 Equipment Requirements
WARNING
THE ELECTRONIC CABINETS (C1 FRONTAL, C1 LATERAL, C2, OPTIONAL
LD CABINET, COOLIX 4100 CHILLERS, DETECTOR CONDITIONERS AND
FLUORO UPS WHEN INSTALLED) INCLUDE FANS THAT ARE CREATING
AIR–CIRCULATION OF PULSED–AIR. WHEN THIS PULSED AIR IS IN AN
ENVIRONMENT THAT MAY CONTAIN AIRBORNE PATHOGENS LIKE AN
EXAM ROOM/CONTROL ROOM, THERE IS A RISK OF TRANSMISSION OF
THESE AIRBORNE PATHOGENS FROM PATIENTS TO OTHER PATIENTS
OR CLINICAL PERSONNEL (NOSOCOMIAL DISEASES).
TO REDUCE THIS RISK, THE ELECTRONIC CABINETS MUST BE
INSTALLED IN A ROOM SEPARATED FROM EXAM ROOM/CONTROL ROOM,
I.E., TECHNICAL ROOM.
Innova
IGS 620, Innova
TM
IGS 630 Preinstallation Manual
TM
Direction 5694389-1-1EN, Revision 2
49