Chapter 4 Room Layouts

1 Room Layout Considerations

1.1 Radiation Protection

Because X–Ray equipment produces radiation, you may need to take special precautions
or make special site modifications. The General Electric Company does not make
recommendations regarding radiation protection. It is the purchasers' responsibility to consult
a radiation physicist for advise on radiation protection in x–ray rooms.

1.2 Service Access

Allow appropriate space for service access of equipment. Consult component pre–installation
directions for clearance information.

1.3 Clinical Access

Make sure that you plan the room with the following clinical access requirements:
Provide easy access to the patient table. Stretchers and other mobile hospital
equipment must reach the table quickly.
The layout of the table in the room (PIM) shall make a provision so that the clearance
between the maximum table position (head side) on system axis and any object in
the room (e.g.: wall, device) be greater than 50 cm (19.7 in)or 65 mm (25.6 in)
if the Header Extender is used), taking into account the fact that the Tilt table can
rotate 180 .
Provide sufficient space around the patient table for the unimpeded conduct of CPR
(Cardiac Pulmonary Resuscitation). With the table in this position, the table must
be capable of rotating +/- 45
Clinicians at the patient table must be able to communicate with assistants in the
control area.
There must be an unrestricted view of the video monitors and physiological monitoring
equipment from the vascular table.
Operators in the control area must have easy access to the control console. However,
position the controls (including handswitches) so that the operator cannot take
exposures while looking around or standing outside the control booth's lead glass
window.
Operators in the control area must have easy access to video recorders and injector
programmers, film and video storage cabinets, and service and operating manuals.
Consult customer on the number and location of nonelectrical lines (air, oxygen,
vacuum, water, etc.) in the vascular room.

1.4 Peripheral Equipment

Consult hospital personnel regarding additional space requirements for the following types of
hospital equipment:
Storage cabinets.
Sinks.