Fault
Yellow service LED is flashing a repeated sequence
Cause
A service is required.
Remedy
See
Flashing service codes
Fault
The red alarm LED is on
Cause
Normal during the software upgrade process. At other times indicates a problem
has been detected in the FLASH memory.
Remedy
If during software upgrade then complete the upgrade process. Otherwise cycle
power and if this does not help then perform a software upgrade.
Fault
The red alarm LED is flashing
Cause
A fail condition has become active.
Remedy
Note the position of the long flashes within the series of 6 flashes to work out the
error code.
Look up the flashing error code in
advice given.
Fault
Blue bearing LED is flashing a repeated sequence
Cause
A bearing service due to inactivity is required.
Remedy
See
Flashing inactivity service codes
Fault
Blue bearing LED is lit
Cause
Bearing run in mode (BRIM) is running.
Remedy
Wait until BRIM is finished. This may take 4.3 hours. We recommend not to cancel
BRIM.
Fault
Pump is not accelerating to full speed
Cause
Bearing run in mode (BRIM) is running.
Remedy
Wait until BRIM is finished. This may take 4.3 hours. We recommend not to cancel
BRIM, refer to
Fault
Any other problems
Cause
Fault conditions other than the ones previously mentioned.
Remedy
Contact the supplier or us.
11.1 Command set error codes
Error codes that may be returned for serial control and monitoring.
Table 26 Command error codes
Returned error code
B8J200880_D
300812671_002_C3
B8J200880_D - Fault finding
on page 100.
Flashing error codes
Bearing run in mode
Meaning
0
No error
Page 99
on page 100 and follow the
on page 101
on page 107.