Occlusion Pressure Limits - CareFusion Alaris PK Plus MK4 8005PK201 Directions For Use Manual

Syringe pump
Table of Contents

Advertisement

Occlusion Pressure Limits

Time to alarm following occlusion is achieved in less than 30 minutes at rates of 1 ml/h and higher by the appropriate selection of
occlusion levels.
The following graphs show the typical values for time to alarm and bolus volume that can be expected in the event of an occlusion when
the BD Plastipak 50 ml syringe is selected with a G40020B standard extension set.
Time to alarm - 1.0 ml/h
Occlusion Level
Bolus Volume
Occlusion Level
Tests at low alarm levels may alarm immediately - the force at these levels is commonly less than the friction in the syringe (with no
additional fluid pressure). The result is that the pressure relating to the low forces will be less than the nominal quoted occlusion
pressure.
— typical
— typical
1000DF00741 Issue 5
Alaris™ PK Plus Syringe Pump
Occlusion Pressure Limits
Time to alarm - 5.0 ml/h
Occlusion Level
43/52
— typical