Service Configuration Mode

This section comprises of a list of options which can be configured. Some can be entered via the pump SERVICE menu (available in
Technician Mode) and others through the Alaris Editor Software.
Enter the access code on the pump for SERVICE mode, then select CONFIGURATION, see the Technical Service Manual for details.
Use Alaris Editor to configure the pump configuration, drug library and units enabled for each data set.
Access codes should only be entered by qualified technical personnel.
A
Date & Time
1. Select DATE & TIME from the CONFIGURATION menu using the
2. Press the OK softkey to confirm.
f
3. Use the
keys to adjust the date displayed, pressing the NEXT softkey to access the next field.
4. When the correct date and time are displayed press the OK softkey to return to the CONFIGURATION menu.
5. Press the QUIT softkey to return to the SERVICE menu and press
Pump Reference
This option is used to add reference text to be shown on the pump start up display.
1. Select PUMP REFERENCE from the CONFIGURATION menu using the
f
2. Use the
keys to enter the text and NEXT to move to the next character.
3. When the desired text has been selected press OK softkey to return to the CONFIGURATION menu.
4. Press QUIT to exit back to the main SERVICE menu and press
Language
This option is used to set the language of messages shown on the pump display.
1. Select LANGUAGE from the CONFIGURATION menu using the
f
2. Use the
keys to select the language.
3. When the desired language has been selected press OK softkey to return to the CONFIGURATION menu.
4. Press QUIT to exit back to the main SERVICE menu and press
Backlight & Contrast
This option is used to set the backlight and contrast on the pump display.
1. Select BACKLIGHT & CONTRAST from the CONFIGURATION menu using the
f
2. Use the
keys to adjust BACKLIGHT, CONTRAST and DIMMING. The contrast of the display will change when scrolling
through the numbers. (Use PARAM to scroll between each option)
3. When the desired value has been reached press the OK softkey, then QUIT to get back to the SERVICE menu and press
and power down.
1000DF00546 Issue 5
Alaris™ GP (Guardrails™) Volumetric Pump
f
keys and press the OK softkey.
a
to exit and power down.
f
keys and press the OK softkey.
a
to exit and power down.
f
keys and press the OK softkey.
a
to exit and power down.
f
24/50
Service Configuration Mode
keys and press the OK softkey.
a
to exit