Configured Options

This section comprises of a list of options which are configurable. Some can be entered via the Pump configuration menu (available in
Technician Mode) and others through the Alaris PK Editor Software.
Enter the access code on Alaris PK Plus Syringe Pump for Configured Options, see the Technical Service Manual for details.
w
Access codes should only be entered by qualified technical personnel.
Use Alaris PK Editor to configure general options, drug library and units enabled for each profile and to configure Syringe Brands and
Models to be enabled.
Clock Set
1. Select CLOCK SET from the Configured Options menu using the
f
2. Use the
keys to adjust the date displayed, pressing the NEXT softkey to access the next field.
3. When the correct time and date are displayed press the OK softkey to return to the Configured Options menu.
Language
This option is used to set the language of messages shown on the Pump display.
1. Select LANGUAGE from the Configured Options menu using the
f
2. Use the
keys to select the language.
3. When the desired language has been selected press SELECT softkey to return to the Configured Options menu.
Contrast
This option is used to set the contrast on the Pump display.
1. Select CONTRAST from the Configured Options menu using the
f
2. Use the
keys to select a contrast ratio value. The contrast of the display will change when scrolling through the numbers.
3. When the desired value has been reached press the OK softkey to return to the Configured Options menu.
Alaris PK Plus Syringe Pump General Options
1. Select GENERAL OPTIONS from the Configured Options menu using the
2. Select the option required to enable/disable or adjust and press the MODIFY softkey.
3. When all the required modifications have been carried out press the QUIT softkey.
4. Either select the next configuration option from the menu or turn the Pump off, returning it to operation as required.
NURSE CALL FITTED
Enables Nurse Call (hardware option).
NURSE CALL INVERT
When enabled, the nurse call output is inverted.
RS232 SELECTED
Sets the Pump's communications to use RS232 (hardware option). The NURSE CALL FITTED option must
be enabled to allow RS232 to function.
Power Down Sequence
Enter the access code on Alaris PK Plus Syringe Pump for alternative Power Down Sequence, see the Technical Service Manual for details.
w
Access codes should only be entered by qualified technical personnel.
ENABLED
When running TCI or TIVA with predictive TCI the Pump may only be powered down by stopping the infusion,
selecting NEW OPERATION from the options menu, confirming the selection and then power down the Pump.
DISABLED
In TCI or TIVA with predictive TCI the Pump may be powered down after putting the Pump on hold.
1000DF00741 Issue 5
f
keys and press the OK softkey.
f
keys and press the OK softkey.
f
keys and press the OK softkey.
f
keys and press the OK softkey.
29/52
Alaris™ PK Plus Syringe Pump
Configured Options