To ensure that this pump remains in good operating condition, it is important to keep it clean and carry out the routine maintenance
procedures described below.
Interval
As per Hospital Policy
Each usage
Before the transfer of the pump to a
new patient and as required
If the pump is dropped, damaged, subjected to excessive moisture or high temperature, immediately take it out
w
of service for examination by a qualified service engineer.
All preventative and corrective maintenance and all such activities shall be performed at a compliant work place in
accordance with the information supplied. CareFusion will not be responsible should any of these actions be
performed outside the instructions or information supplied by CareFusion. For Preventative and Corrective
Maintenance instructions please refer to the Technical Service Manual (TSM).
All servicing should only be performed by a qualified service engineer with reference to the TSM.
Please refer to Technical Service Manual for calibration procedures. The units of measurement used in the
w
calibration procedure are standard SI (The International System of Units) units.
The internal rechargeable battery allows continued operation when the AC power is unavailable, for example during patient transfer or AC
power failure. Mean Time To Battery Empty from fully charged @ 5ml/h & 20°C under normal conditions is 6 hours*. From the battery low
alarm it will take about 2½ hours to 90% charge when reconnected to the AC power supply, whether the pump is in use or not.
The battery is maintenance free, sealed Nickel Metal Hydride and requires no routine servicing. However, to achieve optimum operation,
ensure that the battery is fully recharged after full discharge, before storage, and at regular 3 month intervals during storage.
It is recommended that only a qualified service engineer replaces the battery. For further information regarding the replacement of
batteries refer to the Technical Service Manual.
The battery pack used in this Alaris® Syringe Pump is manufactured by CareFusion and includes a proprietary PCB (printed circuit board)
designed specifically for the Alaris® Syringe Pump, and in conjunction with Alaris® Syringe Pump software, controls battery use, charge
and temperature. Any use of battery packs that are not manufactured by CareFusion in the Alaris® Syringe Pump is at your sole risk, and
CareFusion does not provide any warranty for or endorsement on any battery packs that are not manufactured by CareFusion. CareFusion's
product warranty shall not apply in the event the Alaris® Syringe Pump has suffered damage or premature wear, or malfunctions or
otherwise operates incorrectly, as a result of use with a battery pack that is not manufactured by CareFusion.
*95% lower confidence interval of 5 hours 50 minutes

Maintenance

Routine Maintenance Procedures
Routine Maintenance Procedure
Thoroughly clean external surfaces of the pump before and after prolonged period of storage.
1. Inspect AC power supply plug and cable for damage.
2. Inspect case, keypad and plunger for damage.
3. Check Start up self test operation is correct.
Clean the pump by wiping over with a lint-free cloth lightly dampened with warm water and a
standard disinfectant / detergent solution.
Battery Operation
1000DF00333 Issue 6
31/38