Analog Output Calibration Mode - Teledyne 200E Instruction Manual

Chemiluminescence nitrogen oxides analyzer
Table of Contents

Advertisement

Operating Instructions

6.7.3.2. Analog Output Calibration Mode

The analog outputs can be calibrated automatically or manually. In its default mode, the
instrument is configured for automatic calibration of all channels. Manual calibration should
be used for the 0.1V range or in cases where the outputs must be closely matched to the
characteristics of the recording device. Outputs configured for automatic calibration can be
calibrated as a group or individually. Analog calibration needs to be carried out on first
startup of the analyzer (performed in the factory as part of the configuration process) or
whenever re-calibration is required.
To calibrate the outputs as a group press SETUP-MORE-DIAG-ENTR-NEXT-NEXT and
then the following keys:
Exit at any time
to return to the
main DIAG
menu
If any of the channels have
not been calibrated this
message will read NO.
To automatically calibrate an analog channel, press SETUP-MORE-DIAG-ENTR-NEXT-NEXT
and then the following keys:
62
DIAG
ANALOG I / O CONFIGURATION
PREV
NEXT
DIAG AIO
AOUTS CALIBRATED:
< SET SET>
CAL
DIAG AIO
AUTO CALIBRATING CONC_OUT_1
AUTO CALIBRATING CONC_OUT_2
...
AUTO CALIBRATING TEST_OUTPUT
DIAG AIO
AOUTS CALIBRATED:
< SET SET>
CAL
Model 200E Instruction Manual
ENTR
EXIT
NO
If AutoCal has been
EXIT
channel, the message
for that channel will be
YES
EXIT
turned off for any
similar to:
NOT AUTO CAL
CONC_OUT_1
Exit to return to
the I/O
configuration
menu
044100102 Rev A