6.9.9. Ethernet Port Configuration

The optional Ethernet port (Section 5.7.3) communicates with the analyzer through the
COM2 serial port. Refer to Figure 3-2 and Figure 5-2 for location of this option. The Ethernet
board has two operational modes:
Pass-through mode: This is the normal operation mode in which the board actively
passes data between the RS-232 port and the RJ-45 connector on the analyzer's rear
panel. This enables all remote commands to be passed to the analyzer.
Configuration mode: The board stops passing data and is ready to accept and store
configuration parameters and firmware upgrades. The use of a terminal window or
separate configuration program is needed for these low-level configuration changes.
The iChip configuration utility is available at http://www.teledyne-api.com/software/
and provides a convenient, graphical user interface, which runs only on Microsoft
TM
Windows
operating systems. All commands can also be issued via Terminal window
using the
AT+i
website along with other pertinent information.
The Ethernet board has four LEDs that are visible on the rear panel of the analyzer,
indicating its current operating status (Table 6-18).
LED
LNK (green)
ACT (yellow)
TA (green)
RA (yellow)
As soon as the Ethernet option is enabled in the COMM menu, a new sub-menu INET will
appear. This sub-menu is enabled by default when the analyzer was ordered with built-in
Ethernet and is used to set configuration parameters that enable the Ethernet card to
044100102 Rev A
RANGE = 500.0 PPB
NOX =XXX.X
SETUP
PRIMARY SETUP MENU
MORE
EXIT
SECONDARY SETUP MENU
EXIT
COMMUNICATIONS MENU
COM2
EXIT
Test runs
automatically
protocol. Instructions for

Table 6-18: Ethernet Status Indicators

ON when connection to the LAN is valid.
Flickers on any activity on the LAN.
Flickers when the RS-232 port is transmitting data.
Flickers when the RS-232 port is receiving data.
Operating Instructions
SETUP X.X
COM1 MODE:0
SET>
EDIT
SETUP X.X
COM1 BAUD RATE:19200
<SET
SET>
EDIT
SETUP X.X
COM1 : TEST PORT
<SET
TEST
SETUP X.X
TRANSMITTING TO COM1
<SET
TEST
commands are also available at the
AT+i
Function
EXIT
EXIT
EXIT
EXIT returns to
COMM menu
EXIT
81