Table 6-13: Test Parameters Available For Analog Output A4 - Teledyne 200E Instruction Manual

Chemiluminescence nitrogen oxides analyzer
Table of Contents

Advertisement

Model 200E Instruction Manual
To activate the A4 channel and select a test function, follow this key sequence:
SAMPLE
RANGE = 500.0 PPB
< TST TST > CAL
PRIMARY SETUP MENU
SETUP X.X
CFG DAS RNGE PASS CLK
SECONDARY SETUP MENU
SETUP X.X
COMM VARS
DIAG
DIAG
SIGNAL I / O
NEXT
DIAG
ANALOG OUTPUT
PREV
NEXT

Table 6-13: Test Parameters Available for Analog Output A4

Once a TEST function is selected, the instrument begins to report a signal on the A4 output
and adds TEST= to the list of test functions viewable on the display (just before the TIME
display).
044100102 Rev A
NOX=X.X
SETUP
EXIT returns
to the main
MORE
EXIT
SAMPLE
EXIT
ENTR EXIT
ENTR EXIT
Test Channel
NONE
PMT DETECTOR
Ozone flow
sample flow
SAMPLE PRESSure
rcell pressure
rcell temp
manifold temp
izs temp
conv temp
pmt temp
CHASSIS TEMP
HVPS VOLTAGE
1
This refers to the voltage range of the parameter and
not the output signal of the test channel.
DIAG
PREV
NEXT
display
DIAG TCHN
NEXT
DIAG TCHN
PREV
NEXT
Press PREV or NEXT
to move through the
list of available
parameters
(Table 6-9)
Test parameter range
Test channel is turned off
0-5000 mV
0-1000 cm³/min
0-1000 cm³/min
0-40 in-Hg-A
0-40 in-Hg-A
0-70° C
0-70° C
0-70° C
0-500° C
0-50° C
0-70° C
0-5000 V
Operating Instructions
Continue to press NEXT until ...
TEST CHAN OUTPUT
ENTR
TEST CHANNEL: NONE
ENTR
TEST CHANNEL: PMT DETECTOR
ENTR
Press ENTR to
Press EXIT to
select the displayed
return to the
parameter and to
DIAG menu
activate the test
channel.
1
EXIT
EXIT
EXIT
69