Chemiluminescence Detection; The Photo Multiplier Tube; Figure 10-2: No 2 Conversion Principle - Teledyne 200E Instruction Manual

Chemiluminescence nitrogen oxides analyzer
Table of Contents

Advertisement

Model 200E Instruction Manual
Theory of Operation
heated to a temperature of 315° C. The heated molybdenum reacts with NO
in the sample
2
gas and produces a variety of molybdenum oxides and NO according to Equation 10-4.
+
+
°
xNO
yMo
xNO
M
O
(
at
315
C
)
Eq 10-4
2
y
z
Once the NO
in the sample gas has been converted to NO, it is routed to the reaction cell
2
where it undergoes the chemiluminescence reaction described in Equations 10-1 and 10-2.
Figure 10-2: NO
Conversion Principle
2
By converting the NO
in the sample gas into NO, the analyzer can measure the total NO
2
X
(NO+NO
) content of the sample gas. By switching the NO
converter in and out of the
2
2
sample gas stream every 6 - 10 seconds, the M200E analyzer is able to quasi-continuously
measure both the NO and the total NO
content.
X
The NO
concentration, finally, is not measured but calculated by simply subtracting the
2
known NO content of the sample gas from the known NO
content.
X

10.1.3. Chemiluminescence Detection

10.1.3.1. The Photo Multiplier Tube

The M200E uses a photo-multiplier tube (PMT) to detect the amount of light created by the
NO and O
reaction in the reaction cell.
3
A PMT is typically a vacuum tube containing a variety of specially designed electrodes.
Photons enter the PMT and strike a negatively charged photo cathode causing it to emit
electrons. These electrons are accelerated by an applied high voltage and multiply through
a sequence of such acceleration steps (dynodes) until a useable current signal is generated.
This current increases or decreases with the amount of detected light (Section 10.3.2.2 for
044100102 Rev A
157