Izs Permeation Tubes (Options 53 & 55); Table 5-2: Available Permeation Source Options - Teledyne 200E Instruction Manual

Chemiluminescence nitrogen oxides analyzer
Table of Contents

Advertisement

Optional Hardware and Software
Flow rate of the zero air: If the previous two variables are constant, the permeation
rate of the NO
of zero air produces higher concentrations of NO
In order to keep the permeation rate constant, the IZS enclosure is heated by a PID
(Proportional/Integral/Derivative) temperature control loop to a constant 50° C (10° above
the maximum operating temperature of the instrument). A thermistor measures the actual
temperature and reports it to the CPU for control feedback. The flow rate across the IZS
oven is controlled with a critical flow orifice to 60±5 cm³/min.
Please note, that the IZS option does not contain the actual permeation tube, which needs
to be ordered separately. See next option for description.
Gas flow must be maintained at all times for units with a permeation tube
installed. Insufficient gas flow allows NO
Figure 3-12 shows the internal pneumatic connections for a Model 200E with IZS option
installed. Table 5-1 shows the operational state of the valves associated with the IZS option
during the analyzer's various operational modes. SPAN gas on the zero/span valve option is
equivalent with the permeation tube NO
The state of the IZS valves can also be controlled:
Manually from the analyzer's front panel by using the SIGNAL I/O controls located
under the DIAG Menu (Section 6.8.1),
By activating the instrument's CAL or AutoCal features (Section 7.8),
Remotely by using the external digital control inputs (Section 6.9), or
Remotely through the RS-232/485 serial I/O ports (Section 6.9.10).
5.5.3. IZS Permeation Tubes (Options 53 & 55)
One IZS permeation tube is necessary to operate the IZS option (Option 51). Two different
NO
permeation tubes are available, they are identical in size and shape but are designed to
2
have different permeation rates.
Table 5-2:
Option
OPT 53
OPT 55
Each tube is shipped with a calibration certificate, traceable to a NIST standard, specifying
its actual NO
permeation rate to within ±5%. The calibration is performed at a tube
2
temperature of 50°C and a flow rate of 0.56 liters per minute.
36
into the zero air stream will be constant. Therefore, a lower flow rate
2
CAUTION
severely contaminate the instrument.
2
Available Permeation Source Options
Permeation Rate
(± 25%)
421 ng/min
842 ng/min
Model 200E Instruction Manual
.
2
gas to build up to levels that will
2
gas stream on the IZS option.
Approximate NO
2
Concentration
300 - 500 ppb
600 - 1000 ppb
044100102 Rev A