Troubleshooting & Repair
and sub-assemblies in the analyzer. See the wiring interconnect diagram (document
04504) and interconnect list (document 04496) in Appendix D.

11.1.1. Warning Messages

The most common and/or serious instrument failures will result in a warning message
displayed on the front panel. Table A-2 in Appendix A-3 contains a list of warning
messages, along with their meaning and recommended corrective action.
It should be noted that if more than two or three warning messages occur at the same
time, it is often an indication that some fundamental analyzer sub-system (power supply,
relay board, motherboard) has failed rather than an indication of the specific failures
referenced by the warnings. In this case, a combined-error analysis needs to be performed.
The analyzer will alert the user that a warning is active by displaying the keypad labels
MSG and CLR on the front panel and a text message in the top center line of the display as
shown in this example:
The analyzer will also issue a message to the serial port and cause the red FAULT LED on
the front panel to blink.
To view or clear a warning messages press:
<TST TST> keys replaced with
TEST key. Pressing TEST
deactivates warning messages
until new warning(s) are activated.
If warning messages re-appear,
the cause needs to be found. Do
not repeatedly clear warnings
without corrective action.

Figure 11-1: Viewing and Clearing Warning Messages

11.1.2. Fault Diagnosis with Test Functions

Besides being useful as predictive diagnostic tools, the TEST functions, viewable from the
front panel, can be used to isolate and identify many operational problems when combined
with a thorough understanding of the analyzer's theory of operation (Chapter 10). We
recommend to use the APICOM remote control program to download, graph and archive
TEST data for analysis and long-term monitoring of diagnostic data (
190
SAMPLE
AZERO WARNING
< TST TST > CAL
SAMPLE
SYSTEM RESET
TEST
CAL
SAMPLE
RANGE=500 PPB
< TST TST > CAL
SAMPLE
SYSTEM RESET
< TST TST > CAL
Model 200E Instruction Manual
NOX =123.4
MSG CLR
SETUP
NOX = XXX.X
MSG
CLR
SETUP
NOX = XXX.X
MSG
CLR
SETUP
NOX = XXX.X
CLR
MSG
SETUP
Section 6.11.5
MSG indicates that warning
messages are active.
All Warning messages are hidden,
but MSG button appears
Press CLR to clear the current
warning message.
If more than one warning is
active, the next message will
take its place.
Once the last warning has been
cleared, the analyzer returns to
SAMPLE Mode.
).