To define the SAMPLE or REPORT PERIOD, press SETUP-DAS-EDIT-ENTR and the
following keys:
Use the PREV and NEXT
keys to scroll to the data
channel to be edited.
Press SET> key until you reach REPORT PERIOD (OR SAMPLE PERIOD) ...
Set the number of days
between reports (0-366).
Press keys to set hours
between reports in the format :
HH:MM (max: 23:59). This is a
24 hour clock . PM hours are 13
thru 23, midnight is 00:00.
Example 2:15 PM = 14:15

6.10.2.7. Number of Records

The number of data records in the M200E is limited to a cumulative one million data points
in all channels (one megabyte of space on the disk-on-chip). However, the actual number of
records is also limited by the total number of parameters and channels and other settings in
the iDAS configuration. Every additional data channel, parameter, number of samples
setting etc. will reduce the maximum amount of data points somewhat. In general,
however, the maximum data capacity is divided amongst all channels (max: 20) and
parameters (max: 50 per channel).
The iDAS will check the amount of available data space and prevent the user from specify-
ing too many records at any given point. If, for example, the iDAS memory space can
accommodate 375 more data records, the ENTR key will disappear when trying to specify
more than that number of records. This check for memory space may also make an upload
of an iDAS configuration with APICOM or a Terminal program fail, if the combined number
of records would be exceeded. In this case, it is suggested to either try from the front panel
what the maximum number of records can be or use trial-and-error in designing the iDAS
script or calculate the number of records using the DAS or APICOM manuals. To set the
number of records for one channel from the front panel, press SETUP-DAS-EDIT-ENTR and
the following key sequence.
044100102 Rev A
FROM THE DAS PASSWORD MENU
Edit Data Channel Menu
SETUP X.X
0) CONC:
ATIMER,
PREV NEXT
INS
DEL
EDIT
SETUP X.X
NAME:CONC
<SET
EDIT
PRINT
SET>
SETUP X.X
REPORT PERIOD:000:01:00
<SET
SET>
EDIT
PRINT
SETUP X.X
REPORT PERIODD:DAYS:0
0
0
0
SETUP X.X
REPORT PERIODD:TIME:01:01
0
1
0
0
IIf at any time an illegal entry is selected (e.g., days > 366)
the ENTR key will disappear from the display.
Operating Instructions
2,
900
Exits to the main
Data Acquisition
PRNT
EXIT
menu.
EXIT
EXIT
ENTR
EXIT
ENTR accepts the new string and
ENTR
EXIT
returns to the previous menu.
EXIT ignores the new string and
returns to the previous menu.