Figure 7-1: Gas Supply Setup For Determination Of No - Teledyne 200E Instruction Manual

Chemiluminescence nitrogen oxides analyzer
Table of Contents

Advertisement

Model 200E Instruction Manual
Calibrated
NO
gas
2
(high
concentration)
Zero Air Source

Figure 7-1: Gas Supply Setup for Determination of NO

Step two is to set the expected NO
SAMPLE
< TST TST >
SAMPLE A
LOW
SAMPLE
< TST TST > ZERO
M-P CAL
NOX
M-P CAL
NO
CAL
2
M-P CAL
0
ENTR accepts the expected NO
EXIT ignores the expected NO
044100102 Rev A
Gas Dilution
Calibrator
(with O
VENT
Pump
span gas concentration:
2
RANGE = 500.0 PPB
CAL
RANGE TO CAL: LOW
HIGH
RANGE = 500.0 PPB
CONC
CONCENTRATION MENU
NO
CONV
CONVERTER EFFICIENCY MENU
SET
CO SPAN CONC:
400.0 Conc
0
4
0
0
.0
ENTR EXIT
span
2
concentration value.
span
2
concentration value.
generator
3
option)
Sample
Exhaust
Span
Zero Air
NOX=X.X
SETUP
This step only appears if the
analyzer's reporting range is set
NOX=X.X
for AUTO range mode.
Select LOW and press ENTR.
ENTR
Repeat entire procedure for
HIGH range.
NOX=X.X
EXIT
EXIT
EXIT
The expected NO
Make sure that you specify the actual
concentration value of the span gas. To
change this value to the concentration of the
span gas, enter the number by pressing the
key under each digit until the value is visble.
Calibration Procedures
Source of
SAMPLE Gas
removed
during
procedure
MODEL
200E
Conversion Efficiency
2
span concentration value
2
defaults to 400.0 Conc.
111