Setup - Range Configuration (Rnge); Figure 6-1: Analog Output Connector Key - Teledyne 200E Instruction Manual

Chemiluminescence nitrogen oxides analyzer
Table of Contents

Advertisement

Operating Instructions
In order to compensate for CPU clocks which run fast or slow, there is a variable to speed
up or slow down the clock by a fixed amount every day. To change this variable, press:
SAMPLE
RANGE = 500.0 PPB
< TST TST > CAL
PRIMARY SETUP MENU
SETUP X.X
CFG DAS RNGE PASS CLK
SECONDARY SETUP MENU
SETUP X.X
COMM
VARS
DIAG
SETUP X.X
ENTER VARS PASS: 818
8
1
8
SETUP X.X
0 ) DAS_HOLD_OFF=15.0 Minutes
NEXT
JUMP

6.5. Setup - Range Configuration (RNGE)

The analyzer has four analog output signals, accessible through a connector on the rear
panel.
All of these outputs can be configured either at the factory or by the user for full scale
outputs of 0.1 V, 1V, 5V or 10V. Additionally A1, A2 andA3 may be equipped with optional
0-20 mA current loop drivers and configured for any current output within that range
(e.g. 0-20, 2-20, 4-20, etc.).
Channels A1, A2 and A3 report analog signals that are proportional to the NO
NO
concentrations of the sample gas, respectively. Analog output voltage or current can be
2
scaled independently. Here is one configuration example for independent ranges:
48
NOX =XXX.X
SETUP
MORE
EXIT
EXIT
ENTR
EXIT
EDIT PRNT EXIT
ANALOG OUT
A1
A2
+
-
+

Figure 6-1: Analog Output Connector Key

Model 200E Instruction Manual
SETUPX.X
1 )TPC_ENABLE= ON
PREV
NEXT
JUMP
Continue to press NEXT until ...
SETUP X.X
8) CLOCK_ADJ=0 Sec/Day
PREV
JUMP
SETUP X.X
CLOCK_ADJ:0 Sec/Day
+
0
0
Enter sign and number of seconds per
day the clock gains (-) or loses (+).
SETUP X.X
8) CLOCK_ADJ=0 Sec/Day
PREV NEXT JUMP
A3
A4
-
+
-
+
-
EDIT PRNT EXIT
EDIT
PRNT EXIT
ENTR
EXIT
EDIT PRNT
EXIT
3x EXIT returns
to the main SAMPLE display
, NO and
x
044100102 Rev A