Compact Report; Starting Date; Disabling/Enabling Data Channels - Teledyne 200E Instruction Manual

Chemiluminescence nitrogen oxides analyzer
Table of Contents

Advertisement

Model 200E Instruction Manual
To enable automatic COM port reporting, press SETUP-DAS-ENTR and then:
Toggle key to turn
reporting ON or OFF

6.10.2.9. Compact Report

When enabled, this option avoids unnecessary line breaks on all RS-232 reports. Instead of
reporting each parameter in one channel on a separate line, up to five parameters are
reported in one line, instead. For example, channel DIAG would report its record in two lines
(10 parameters) instead of 10 lines. Individual lines carry the same time stamp and are
labeled in sequence.

6.10.2.10. Starting Date

This option allows to specify a starting date for any given channel in case the user wants to
start data acquisition only after a certain time and date. If the Starting Date is in the past,
the iDAS ignores this setting.

6.10.2.11. Disabling/Enabling Data Channels

Data channels can be temporarily disabled, which can reduce the read/write wear on the
disk-on-chip. The HIRES channel of the M200E, for example, is disabled by default.
044100102 Rev A
FROM THE DAS PASSWORD MENU
Edit Data Channel Menu
SETUP X.X
0) CONC:
ATIMER,
PREV NEXT
INS
DEL
SETUP X.X
NAME:CONC
<SET
EDIT
PRINT
SET>
Press SET> key until...
SETUP X.X
RS-232 REPORT: OFF
<SET
SET>
EDIT
PRINT
SETUP X.X
RS-232 REPORT: OFF
OFF
Operating Instructions
2,
900
EDIT
PRNT
EXIT
EXIT
EXIT
ENTR
EXIT
Exits to the main
Data Acquisition
menu
ENTR accepts the new
setting and returns to the
previous menu.
EXIT ignores the new setting
and returns to the previous
menu.