Figure 3-6: Basic Pneumatic Connections For Units With Valve Options - Teledyne 200E Instruction Manual

Chemiluminescence nitrogen oxides analyzer
Table of Contents

Advertisement

Getting Started
MODEL 700
Gas Dilution Calibrator
(with O
generator option)
3
MODEL 701
Zero Air
Calibrated
Generator
NO gas
(high
concentration)

Figure 3-6: Basic Pneumatic Connections for Units With Valve Options

Once the appropriate pneumatic connections have been made, check all pneumatic fittings
for leaks using the procedures defined in Section 11.5.
Gas flow must be maintained at all times for units with IZS Options installed. The
IZS option includes a permeation tube which emits NO
build up NO
to levels that will damage the instrument. Remove the permeation
2
device when taking the analyzer out of operation.
16
Source of
SAMPLE Gas
VENT
External Zero
Air Scrubber
Source of
SAMPLE Gas
Pump
External Zero
Air Scrubber
Particulate
Filter
WARNING
Model 200E Instruction Manual
Zero/Span Valves – Option 50
VENT if input is pressurized
Sample
Exhaust
Pump
Span
Zero Air
Filter
Internal Zero/Span Option (IZS) – Option 51
VENT if input is pressurized
Sample
MODEL
Exhaust
200E
Span
Zero Air
. Insufficient gas flow can
2
MODEL
200E
044100102 Rev A