First Calibration; Basic Calibration Procedure - Teledyne 200E Instruction Manual

Chemiluminescence nitrogen oxides analyzer
Table of Contents

Advertisement

Getting Started

3.6. First Calibration

3.6.1. Basic Calibration Procedure

The initial calibration should be carried out using the 500 ppb range, a span gas with about
400 ppb NO and with the unit set for SINGLE range, which will enable you to compare your
calibration results to the factory calibration as listed on the Final Test and Validation Data
Sheet. The following three-step procedure assumes that the instrument does not have any
of the available zero/span (Z/S) valve options installed. Chapter 7 contains instructions for
calibrating instruments with Z/S valve options, Chapter 8 for EPA protocol calibration.
Set the analog output range of the M200E:
Press this button to set
the analyzer for SNGL
DUAL or IND ranges
To change the value of the
reporting range span, enter the
number by pressing the key under
each digit until the expected value
appears.
Set the expected NO and NOx span gas concentration. If you supply NO gas to the
analyzer, the values for expected NO and NO
The NO
& NO span concentration
x
values automatically default to
400.0 Conc.
To change this value to the actual
concentration of the span gas, enter
the number by pressing the key
under each digit until the expected
value appears.
26
SAMPLE
RANGE = 500.0 PPB
< TST TST > CAL
SETUP X.X
CFG DAS
RNGE
PASS CLK MORE
SETUP X.X
RANGE CONTROL MENU
MODE SET UNIT
SETUP X.X
RANGE: 500.0 CONC
0
0
5
0
0
.0
SETUP X.X
RANGE:
500.0 Conc
0
0
0
5
0
.0
need to be identical.
x
SAMPLE
RANGE = 500.0 PPB
< TST TST >
CAL
M-P CAL
RANGE = 500.0 PPB
< TST TST >
ZERO
CONC
M-P CAL
CONCENTRATION MENU
NOX
NO
CONV
M-P CAL
NOX SPAN CONC: 400.0 Conc
4
0
.0
0
0
0
M-P CAL
CONCENTRATION MENU
NOX
NO
CONV
M-P CAL
NO SPAN CONC: 400.0 Conc
0
0
0
4
0
.0
Model 200E Instruction Manual
NOX= X.X
SETUP
EXIT
EXIT
Press this button to select the
concentration units of measure:
PPB, PPM, UGM, MGM
ENTR EXIT
EXIT ignores the new setting and
returns to the RANGE CONTROL
MENU.
ENTR accepts the new setting and
returns to the
RANGE CONTROL MENU.
ENTR EXIT
NOX= X.X
SETUP
This sequence causes the
analyzer to prompt for the
expected NO and NO
span concentration.
NO = X.X
EXIT
EXIT
ENTR EXIT
EXIT ignores the new setting and
returns to the CONCENTRATION
ENTR accepts the new setting and
EXIT
returns to the
CONCENTRATION MENU..
ENTR EXIT
044100102 Rev A
x
MENU.