Internal Zero/Span (Izs) (Option 51); Table 5-1: Izs Or Z/S Valve States - Teledyne 200E Instruction Manual

Chemiluminescence nitrogen oxides analyzer
Table of Contents

Advertisement

Model 200E Instruction Manual
Mode
Valve
SAMPLE
Sample/Cal
Zero/Span
ZERO CAL
Sample/Cal
Zero/Span
SPAN CAL
Sample/Cal
Zero/Span
The state of the IZS valves can also be controlled:
Manually from the analyzer's front panel by using the SIGNAL I/O controls located
under the DIAG Menu (Section 6.8.1),
By activating the instrument's AutoCal feature (Section 7.8),
Remotely by using the external digital control inputs (Section 6.9), or
Remotely through the RS-232/485 serial I/O ports (Section 6.9.10).
Sources of zero and span gas flow must be capable of supplying at least 600 cm
the analyzer draws about 500±50 cm³/min into the instrument. Both supply lines should be
vented outside of the analyzer's enclosure. In order to prevent back-diffusion and pressure
drop effects, these vent lines should be between 2 and 10 meters in length.

5.5.2. Internal Zero/Span (IZS) (Option 51)

The M200E can be equipped with an internal zero air and span gas generator system (IZS).
This option includes a heated enclosure for a NO
producing zero air and two valves for switching between the sample gas inlet and the
output of the zero/span subsystem, functionally identical to the zero/span valve option.
The IZS option includes a two stage, external zero air scrubber assembly that removes all
NO & NO
from the zero air source. The scrubber is filled with 50% Purafil Chemisorbant
2
(for conversion of NO to NO
assembly also includes a small particle filter.
Span gas is created by passing zero air over a NO
liquid NO
, which slowly permeates through a permeable membrane into the surrounding
2
air. The concentration of the NO
Size of the membrane: The larger the area of the membrane, the more permeation
occurs.
Temperature of the NO
increases the pressure inside the tube and therefore increases the rate of
permeation.
044100102 Rev A
Table 5-1:
IZS or Z/S Valve States
Condition
Open to sample gas inlet
Open to zero air inlet
Open to zero/span inlet (activated)
Open to zero air inlet
Open to zero/span inlet (activated)
Open to span gas inlet / IZS gas (activated)
) and 50% activated charcoal (for removal of NO
2
span gas is determined by three factors:
2
: Increasing the temperature of the permeation tube
2
Optional Hardware and Software
permeation tube, an external scrubber for
2
permeation tube. This tube contains
2
Valve Port Connection
NO = normally open port
NC = normally closed port
COM = common port
NO ! COM
NO ! COM
NC ! COM
NO ! COM
NC ! COM
NC ! COM
3
/min as
®
). This
2
35