3.5.4. Functional Check

After the analyzer's components have warmed up for at least 30 minutes, verify that the
software properly supports any hardware options that were installed.
Check to make sure that the analyzer is functioning within allowable operating parameters.
Appendix A and C include a list of test functions viewable from the analyzer's front panel as
well as their expected values. These functions are also useful tools for diagnosing perform-
ance problems with your analyzer (Chapter 11). The enclosed Final Test and Validation Data
Sheet (part number 04490) lists these values before the instrument left the factory. To
view the current values of these test functions press the <TST TST> keys:
Figure 3-11 and Figure 3-12 show the analyzer flow diagrams without and with options
installed. Refer to these diagrams whenever troubleshooting or a thorough understanding of
the analyzer operation is required. Note that valve options for zero/span valves or IZS are
principally the same, except that the span gas stream is connected differently in the
zero/span option.
More detailed pneumatic diagrams for all configurations (base, zero/span valves, IZS,
sample dryer) can be found in Appendix D, diagram number 04574.
044100102 Rev A
SAMPLE
RANGE = 500.0 PPB
< TST TST >
CAL
Toggle <TST TST> keys to
scroll through list of functions
1
Only appears if IZS option is
installed.
2
Only appears if analyzer is set
to display NO, NO
, NO
2
x
simultaneously in message
field.
3
Only appears if analog autput
A4 is actively reporting a test
function
NOX = X.X
SETUP
RANGE
NOX STB
SAMP FLW
OZONE FL
PMT
NORM PMT
AZERO
HVPS
RCELL TEMP
Refer to
BOX TEMP
Section
PMT TEM
6.2.1 for
1
IZS TEMP
definitions
MOLY TEMP
of these
RCEL
test
SAMP
functions.
NOX SLOPE
NOX OFFS
NO SLOPE
NO OFFS
2
NO2
2
NOX
2
NO
3
TEST
TIME
Getting Started