6.7.7. Ozone Generator Override

This feature allows the user to manually turn the ozone generator off and on. This can be
done before disconnecting the generator, to prevent ozone from leaking out, or after a
system restart if the user does not want to wait for 30 minutes during warm-up time. Note
that this is one of the two settings in the DIAG menu that is retained after you exit the
menu. To access this feature press the following keys. Also note that the ozone generator
does not turn on if the ozone flow conditions are out of specification (e.g., if there is no flow
through the system or the pump is broken).
SAMPLE
RANGE = 500.0 PPB
< TST TST > CAL
SETUP X.X
PRIMARY SETUP MENU
CFG DAS RNGE PASS CLK
SETUP X.X
SECONDARY SETUP MENU
COMM VARS
DIAG
SETUP X.X
ENTER DIAG PASS: 818
8
1
8
044100102 Rev A
NOX=X.X
SETUP
MORE
EXIT
EXIT
ENTER DIAG PASS: 818
ENTR
EXIT
While the electrical test is
activated, PMT should equal:
2000 mV ± 1000 mV
NOX=X.X
SETUP
MORE
EXIT
EXIT
ENTR
EXIT
DIAG
SIGNAL I / O
PREV
NEXT
JUMP
Press NEXT until...
DIAG
ELECTRICAL TEST
PREV NEXT
DIAG ELEC
RANGE = 500.0 PPB
<TST TST>
Press TST until...
DIAG ELEC
PMT = 1732 MV
<TST TST>
DIAG
SIGNAL I / O
PREV NEXT
JUMP
ENTR
Press NEXT until...
DIAG
OZONE GEN OVERRIDE
PREV NEXT
DIAG OZONE
OZONE GEN OVERRIDE
OFF
Operating Instructions
ENTR
EXIT
ENTR
EXIT
NOX=X.X
EXIT
NOX=X.X
EXIT
EXIT
ENTR
EXIT
Toggle this key to
turn the O
3
ENTR
EXIT
generator ON/OFF.