Control Inputs; Figure 6-4: Status Output Connector; Table 6-14: Status Output Pin Assignments - Teledyne 200E Instruction Manual

Chemiluminescence nitrogen oxides analyzer
Table of Contents

Advertisement

Model 200E Instruction Manual
The status outputs are accessed through a 12 pin connector on the analyzer's rear panel
labeled STATUS. The function of each pin is defined in Table 6-14.
1
Connector
Pin
1
SYSTEM OK
2
CONC VALID
3
HIGH RANGE
4
ZERO CAL
5
SPAN CAL
6
DIAG MODE
7
LOW RANGE
8
D
EMITTER BUS
+
DC POWER
DIGITAL
GROUND

6.8.2. Control Inputs

Control inputs allow the user to remotely initiate ZERO and SPAN calibration modes are
provided through a 10-pin connector labeled CONTROL IN on the analyzer's rear panel.
These are opto-isolated, digital inputs that are activated when a 5 VDC signal from the "U"
pin is connected to the respective input pin.
044100102 Rev A
STATUS
2
3
4
5

Figure 6-4: Status Output Connector

Table 6-14: Status Output Pin Assignments

Status
ON if no faults are present.
ON if concentration measurement is valid, OFF when invalid.
ON if unit is in high range of any AUTO range mode.
ON whenever the instrument is in ZERO calibration mode.
ON whenever the instrument is in SPAN calibration mode.
ON whenever the instrument is in DIAGNOSTIC mode.
ON if unit is in low range of any AUTO range mode.
Unused.
The emitters of the transistors on pins 1-8 are bussed
together. For most applications, this pin should be connected
to the circuit ground of the receiving device.
+ 5 VDC, 30 mA maximum (combined rating with Control
Inputs).
The ground from the analyzer's internal, 5/±15 VDC power
supply.
6
7
8
D
Condition (ON=conducting)
Operating Instructions
+
73