Sample Period And Report Period - Teledyne 200E Instruction Manual

Chemiluminescence nitrogen oxides analyzer
Table of Contents

Advertisement

Model 200E Instruction Manual
To configure the parameters for a specific data parameter, press:
FROM THE EDIT DATA PARAMETER MENU
(see previous section)
SETUP X.X
0) PARAM=PMTDET, MODE=INST
PREV NEXT
INS
SETUP X.X
PARAMETERS: PMTDET
SET> EDIT
SETUP X.X
SAMPLE MODE: INST
<SET SET>
EDIT
SETUP X.X
PRECISION: 1
<SET SET>
EDIT
SETUP X.X
STORE NUM. SAMPLES: OFF
<SET
EDIT

6.10.2.6. Sample Period and Report Period

The iDAS defines two principal time periods, the SAMPLE PERIOD, which is the time
interval at which data are sampled (in volatile memory; not stored permanently), and the
REPORT PERIOD, which is the time interval of which data are permanently stored to the
Disk-on-Chip data storage and reported to the communication ports.
044100102 Rev A
DEL
EXIT
EDIT
EXIT
SETUP X.X
PARAMETERS: PMTDET
PREV NEXT
Cycle through list of available
Parameters.
EXIT
SETUP X.X
SAMPLE MODE: INST
INST
AVG
MIN
MAX
Press the key for the desired mode
EXIT
SETUP X.X
PRECISION: 1
1
Set for 0-4
EXIT
SETUP X.X
STORE NUM. SAMPLES: OFF
OFF
Turn ON or OFF
Operating Instructions
ENTR
EXIT
EXIT
EXIT
ENTR
EXIT
ENTR accepts the new
setting and returns to the
previous menu.
EXIT ignores the new setting
and returns to the previous
<SET Returns to
previous
Functions
95

Hide quick links:

Advertisement

Table of Contents

Troubleshooting

loading

Table of Contents