Malfunctions and Troubleshooting
Display
Error 312
TS-X BLOCKED
Error 313
TS-Z BLOCKED
Error 314
TS-AX BLOCKED
Error 315
TS-AY BLOCKED
92
Cause
The transfer arm of the
TS5025/5015 Transfer
Station is blocked in its left/
right movement (X-axis).
The transfer arm of the
TS5025/5015 Transfer
Station is blocked in its up/
down movement (Z-axis).
The gripper of the TS5015
Transfer Station is blocked
in its left/right movement
(X-axis).
The carriage of the TS5015
Transfer Station is blocked
in its forward/backward
movement (Y-axis).
Troubleshooting
Remove the cause of the blockage (specimen
slide holder) and inspect the transfer arm of
the TS5025/TS5015. Remove empty specimen
slide holders from the discharge chute of
the transfer station. Switch the instrument
off and on. Proceed further as described in
(→ p. 42 – 5.3 Switching the instrument
on or
off). Place any removed specimen slide
holders into the coverslipper and have them
coverslipped (thus outside of the workstation
operation).
Remove the cause of the blockage (specimen
slide holder) and inspect the transfer arm of
the TS5025/TS5015. Remove empty specimen
slide holders from the discharge chute of
the transfer station. Switch the instrument
off and on. Proceed further as described in
(→ p. 42 – 5.3 Switching the instrument
on or
off). Place any removed specimen slide
holders into the coverslipper and have them
coverslipped (thus outside of the workstation
operation).
Remove the cause of the blockage (specimen
slide holder) and inspect the gripper of the
TS5015. Remove empty specimen slide holders
from the discharge chute of the transfer
station. Switch the instrument off and on.
Proceed further as described in
5.3 Switching the instrument on or
any removed specimen slide holders into the
coverslipper and have them coverslipped (thus
outside of the workstation operation).
Remove the cause of the blockage (specimen
slide holder) and inspect the carriage of the
TS5015. Remove empty specimen slide holders
from the discharge chute of the transfer
station. Switch the instrument off and on.
Proceed further as described in
5.3 Switching the instrument on or
any removed specimen slide holders into the
coverslipper and have them coverslipped (thus
outside of the workstation operation).
(→ p. 42 –
off). Place
(→ p. 42 –
off). Place
Version 3.4, Revision M