Leica CV5030 Instructions For Use Manual page 55

Robotic cover slipper
Hide thumbs Also See for CV5030:
Table of Contents

Advertisement

Display
Leica CV5030
Explanation
The number of cover slips
has fallen below the limit.
The instrument continues to
work until the PAUSE button
is pressed or all cover slips
have been used up.
The instrument is in pause
mode. The coverslipping
process was interrupted
due to the empty cover slip
magazine.
The instrument is in pause
mode. There is no loading
bath in the loading drawer.
The instrument is in pause
mode since a vacuum could
not be attained in the system.
The instrument is in pause
mode; the working pressure
in the system could not be
attained.
The instrument is in pause
mode; the system was unable
to grab a specimen slide.
At least three cover slips in a
row are defective.
Operation
Action
Press PAUSE to interrupt
coverslipping and refill the cover
slip magazine. Press START to
continue working.
Fill the cover slip magazine.
Pressing START allows work to
continue.
Insert a loading bath into the
loading drawer. Press RESPOND as
confirmation, then press START to
continue coverslipping.
Check the coverslip mountant
bottle/dispenser system for leaks
and whether both O-rings are
present; replace the O-rings, if
necessary.
Check the coverslip mountant
bottle/dispenser system for leaks
and whether both O-rings are
present; replace the O-rings, if
necessary.
In the specimen slide holder, check
whether the specimen slides are
inserted correctly and undamaged.
If the message occurs repeatedly
despite the specimen slides being
inserted correctly, remove the
specimen slides and apply cover
slips by hand.
Check the cover slips in the cover
slip magazine. If the messages
occurs repeatedly despite the cover
slips being intact, clean the cover
slip sensor. If the message persists,
clean the Pick & Place module
(skids and suction cups).