Fig. 38
5.9

Setting parameter sets

L There are a total of four programs
different parameter settings.
• Pressing the START button calls up the next higher program.
• Pressing the PAUSE button calls up the next lower program (→ Fig. 39).
Fig. 39
5.10 MENU A — parameter settings
L The individual buttons used to select and program the respective parameters are in the bottom
row of the control panel field.
Leica CV5030
MENU B
includes the following parameters:
1
CSP
- Cover slip placement position
STP
- Starting point for the coverslip mountant
stroke
SPK
- Volume of the signal tone
LIM
- Cover slip limit setting
(SET 1
to
SET
4) that can be stored in the instrument with
5
Operation
59