Fig. 50
Important instructions for operation of different specimen slide holders
Note
The default specimen slide holders intended for workstation operation are Leica plastic holders
for 30 specimen slides (Order No. 14 0475 33643). Use of other Leica specimen slide holders (e.g.
Leica plastic holders for 20 specimen slides) or specimen slide holders from other manufacturers
is not permissible in workstation operation. Specimen slide holders from other manufacturers are
only permissible when they have been tested on the instrument. These specimen slide holders, or
compatible bath inserts, are listed in
slide holders may in some cases not be recognized by the instrument and a collision may result.
• Apart from workstation operation, the robotic coverslipper can be manually loaded with specimen slide
holders from other manufacturers, or with other Leica specimen slide holders, with use of a compatible
bath insert only.
• After coverslipping is concluded, ensure that the specimen slide holder is removed once again.
Likewise, the replaced bath insert is to be removed before starting a new workstation operation.
Leica CV5030
1
(→ p. 99 – 9. Optional
Workstation operation
Accessories). Impermissible specimen