Advertisement

Warning
IMPORTANT!
BEFORE switching off the instrument, remove the loading bath, cover it and place it under a laboratory fume
hood until the next time the instrument is started. Only reinsert the loading bath AFTER the instrument is
switched back on and initialized to avoid changes to settings and damage to the gripper.
Warning
IMPORTANT!
In general, when inserting or removing the dispenser, make sure that caution is used during movements at
the time of insertion or removal. If the dispenser falls down, it can cause injuries to users and damage to the
dispenser group/instrument.
Leica CV5030
When the loading drawer with loading bath is pulled out,
BATH NOT READY
appears on the display.
A loaded specimen slide holder can now be inserted into the
loading bath. After inserting it, push the loading drawer back
and close the loading door.
The message
PRIME POSITION
out of the rest position; the message
displayed.
Press the PRIME button once. The message
appears and the instrument continues to generate pressure
in the coverslip mountant bottle until the pump noise can no
longer be heard and the message
Now press the PRIME button a second time. Hold the button
until the coverslip mountant comes out of the dispenser
without bubbles. Do not dispense coverslip mountant into
the existing glass vial; use an alternate container instead
(e.g. upper part of the cover slip packaging).
If sufficient coverslip mountant has been dispensed, the
message
DISP. POSITION
into working position. The message disappears and the
instrument is ready to operate. The displays shows the
message READY.
If the READY LED switches from red to green, the
coverslipping operation can be started by pressing the
START button.
Operation
appears. Take the dispenser
PRIME TO ACT.
ACTIVATING
PRIME
appears.
appears. Move the dispenser
5
is
43