Note
In order to be able to degas air bubbles generated while filling the mountant bottle (→ Fig. 30-3), the
coverslip mountant must sit for 6 - 12 hours (the period depends on the type of coverslip mountant).
• After filling the mountant bottle (→ Fig. 30-3), check the bottle neck and threading for cover slip
mountant residue and clean it if necessary.
• To achieve optimal coverslipping quality, leave the filled mountant bottle under a fume hood for 6 - 12
hours covered by a loosely positioned, blue bottle lid. Only insert the bottle into the instrument after
doing this.
• Screw the red lid
so, ensure that the black O-ring is present and positioned correctly on the dispensing group and the
mountant bottle's blue shutter ring.
Warning
Only glass bottles provided by Leica may be used (see
(→ p. 17 – 3.3 Standard delivery – packing
not be used on the instrument.
Leica CV5030
(→ Fig. 30-4)
for the dispenser group tightly onto the mountant bottle. While doing
list). Other glass or plastic bottles are not suitable and must
Instrument Setup
(→ p. 99 – 9. Optional Accessories)
4
and
39