5
Operation
The pressure needed to dispense the coverslip mountant is set in increments of 1:
1 = 100 mbar/lowest pressure to 10 = 1000 mbar/highest pressure
• An increase of "1" increases the pressure by 100 mbar.
• The pressure has to be adapted based on the coverslip mountant used (different viscosities) and the
dispenser needle used (different diameters).
Stroke Length Corr — stroke length correction of the coverslip mountant
The stroke length for the coverslip mountant set by the
parameter can be shortened by changing
stroke).
The reduction only applies from the label field (→ Fig. 42-1)
• The factory preset value is 0.
In increments of 10 from:
0 = no change to -100 = maximum change
Note
The application quantity is not changed, but only the length of the coverslip mountant stroke.
2
Fig. 42
62
LEN
(→ Fig. 42-2)
LEN
=
STROKE LENGTH CORRECTION
1
(length correction for the coverslip mountant application)
STK
= MOUNTANT STROKE
(length of the coverslip mountant application)
STP
= STROKE POSITION CORRECTION
(starting point of the coverslip mountant application)
CSP
= COVERSLIP POSITION CORRECTION
(cover slip placement position)
STK
(length of the coverslip mountant stroke)
(length correction for the coverslip mountant
area to the middle of the specimen slide.