Note
If the RESPOND button is not pressed immediately, the signal tone stops automatically after approx.
15 seconds and the READY LED is illuminated in red.
The RESPOND button has to be pressed so that the READY LED is illuminated in green and work on
the instrument can continue.
Leica CV5030
• The message FINISHED has to be confirmed with the
RESPOND button. Afterwards, the READY LED is
illuminated in green and the message
the display.
• The output magazine stays at the position of the
specimen slide inserted last.
• Press the LIFT button to remove the output magazine. The
output magazine moves to the output position.
• The output magazines have to be removed and unloaded
manually.
• Afterwards, place the empty output magazines in the
output station.
• Remove the empty specimen slide holder from the loading
bath (not necessary in workstation mode) and replace it
with a full specimen slide holder.
• Press the START button to continue coverslipping.
• At the end of the workday and before long work breaks,
move the dispenser into the park position to keep the
dispenser needle from drying out.
• If some or all specimen slides are to be removed, use the
LIFT key to raise the output magazines to the removal
position. Removal is not absolutely necessary after each
specimen slide holder, as long as sufficient capacity
remains in the output magazine for the next coverslipping
cycle (max. capacity = 60 specimen slides).