Output magazine full
The specimen slide cannot be gripped
Leica CV5030
• Then, continue the coverslipping operation by pressing
START.
The coverslipping operation is interrupted if all of the
positions in the output magazine are filled by specimen
slides. The instrument is in PAUSE mode.
• The READY LED is illuminated in red and PAUSE is green.
The display shows
PAUSED CHECK
by a signal tone that shows that the output magazine is
missing or has to be emptied.
• Confirm the signal tone with RESPOND.
• The instrument shows
• Remove the full output magazines and insert empty
magazines.
• Press START and the instrument continues working.
If the gripper fails to grab hold of a specimen slide after
three attempts, the device automatically switches to
mode.
• The READY LED is illuminated in red, the PAUSE
LED is green. The display shows
SLIDES, accompanied by a signal tone that shows that
coverslipping cannot continue.
• Confirm and switch off the signal tone with RESPOND.
• The instrument shows
• Open the load drawer for the loading bath and, inside the
specimen slide holder, check:
1. Whether the individual specimen slides are inserted
correctly.
2. Whether the distances between the individual
specimen slides are correct.
3. Whether the specimen slide is damaged.
• Consider moving the ungripped specimen slide forward by
one position in the specimen slide holder or removing it
and applying a cover slip by hand.
Operation
LIFT, accompanied
PAUSED
on the display.
PAUSED CHECK
PAUSED
on the display.