Leica CV5030 Instructions For Use Manual page 61

Robotic cover slipper
Hide thumbs Also See for CV5030:
Table of Contents

Advertisement

Fig. 41
Mountant Volume — volume of the coverslip mountant
The opening time for the dispenser valve is set in increments of 1:
Adjustment options for the opening time:
In increments of 1 from 1 (short) to 9 (long).
1 = less coverslip mountant to 9 = maximum quantity of coverslip mountant
• The quantity of the coverslip mountant applied also depends on the composition of the mountant (e.g.
viscosity) and the selected dispenser needle (different diameters).
Mountant Type — dispensing pressure
Leica CV5030
Label field (→ Fig. 41-1),
1
Mountant Stroke: 40 = 40 mm Cover slip length
Mountant Stroke: 50 = 50 mm Cover slip length
Mountant Stroke: 55 = 55 mm Cover slip length
Mountant Stroke: 60 = 60 mm Cover slip length
2
Stroke lenght
(→ Fig. 41-2)
5
Operation
61