6
Workstation operation
6.4

Interrupting workstation operation

Warning
If the instrument is operated as a workstation, the user must ensure that before reinitialization, no rack is in
the transfer station.
Note
So that you do not disturb workstation operation, you should process specimen slide holders that
come from the slide stainer first. Only then should you process the specimen slide holders to be loaded
manually in the coverslipper.
• In case of power failure or if the instruments are switched off, specimen slide holders can remain in
different positions. Therefore, if this error occurs, the following items must be checked:
a. Transfer station with all transfer facilities
b. Loading bath of the coverslipper
• These specimen slide holders must be removed manually to prevent collisions after resuming
workstation operation.
• The specimen slide holders that are left behind must be placed in the coverslipper manually for further
processing and removed after processing.
80
• The PAUSE and STOP buttons have mostly the same functions outlined in
(→ p. 48 – 5.6 Interrupting the coverslipping
• If the coverslipping operation is interrupted using the STOP button, the
coverslipper must be initialized. To do so, press the START button.
• After the coverslipping operation is interrupted using the STOP button, the
specimen slide holder, which is still in the instrument, must be removed
from the loading bath manually after processing is finished. In this case, it
is not transported automatically back to the transfer station.
operation).
Version 3.4, Revision M