Leica CV5030 Instructions For Use Manual page 54

Robotic cover slipper
Hide thumbs Also See for CV5030:
Table of Contents

Advertisement

5
Operation
Display
54
Explanation
The user is prompted to
check the loading bath.
The loading drawer has been
pulled out.
The user is prompted to move
the dispenser into the park
position.
The user is prompted to
press PRIME.
Dispensing pressure is built
up in the coverslip mountant
bottle.
The user is prompted to press
the PRIME button again in
order to pump mountant from
the dispenser.
The user is prompted to move
the dispenser into working
position.
The instrument is in a state
ready to operate.
PAUSE has been pressed.
A coverslipping operation
for a specimen slide started
previously is concluded.
The instrument is in Pause
mode.
A signal tone is audible.
The instrument is in pause
mode. No output magazine
is present or all output
magazines are full.
Action
Open the loading door, pull out the
loading drawer and loading bath.
Then check whether there is a lid on
the loading bath or an empty rack
inside it (remove both if they are
present).
Push in the drawer and close the
loading door.
Move the dispenser into the prime
position.
Press the PRIME button.
Wait until dispensing pressure
has been reached in the coverslip
mountant bottle and the pump
noise is no longer audible.
Hold the PRIME button down
until mountant comes out of the
dispenser without bubbles. Check
whether there are still air bubbles
in the dispensing group's coverslip
mountant hose.
If air bubbles are still present,
press PRIME until they have been
removed.
Insert the dispenser into the holder
of the working position.
Pressing START begins the
coverslipping operation.
Wait until the instrument is in pause
mode.
Confirm the signal tone with
RESPOND. Press START to
continue the coverslipping
operation.
Check the output position. Empty or
insert new output magazines. Press
START to continue working.
Version 3.4, Revision M