Instrument Setup

4.2.1 Setting up the Leica CV5030

Setting up the Leica CV5030
1. Grab hold of bottom of the instrument from the front and rear (with at least 2 people; the
instrument weighs approx. 57 kg) and place it on a stable laboratory table.
2. When doing so, ensure that the instrument is standing on all four feet.
3. Pull the plastic protective hood off of the instrument by pulling upwards and remove the two
adhesive strips (→ Fig. 5-1).
4. Remove the foam safeguard
5. Open the loading door
6. Ensure that the provided accessories are complete in accordance with the order.
2
Fig. 5
4.3

Preparing and adjusting the instrument

To commission the instrument, carry out the following tasks described in the chapters below:
1. Remove the transport anchors.
2. Insert the filter and attach the exhaust hose.
3. Level the instrument.
4. Install the dispenser group.
5. Align the dispenser needle with the specimen slide outfeed.
6. Install the dispenser needle cleaner.
22
(→ Fig. 5-2)
from the output station.
(→ Fig. 5-3)
and remove the foam cover for the loading bath.
3
1