6
Workstation operation
Warning
Basically, we do not recommend manual loading of the Leica CV5030 in workstation operation, as this can
impair or stop the program flow of the Leica ST5020 Multistainer.
• If it is necessary to insert a specimen slide holder into the loading bath manually for coverslipping,
the empty specimen slide holder must also be removed from the loading drawer manually. The empty
specimen slide holder is not conveyed automatically into the transfer station and thus onto the
discharge chute.
6.3

Important instructions for operation as workstation

Note
• Specimen slide holders loaded manually into the Leica CV5030 must always be removed
immediately after processing.
• The unloading chute
message appears. At the latest, this should be done after a signal tone that occurs after the third
specimen slide holder and the
• Ensure that the coverslip magazine in the Leica CV5030 is filled regularly. This will prevent
complete emptying of the coverslip magazine, which would cause the robotic coverslipper to stop
working and go into pause mode.
• Full output magazines must be removed regularly, as otherwise continuous workstation operation is
not guaranteed.
1
Fig. 49
• The deep bath, which can be identified by side notches (→ p. 103 – Fig. 66), can be used in
workstation mode in combination with a silver/blue transfer arm
is in the rear part of the TS5015 / TS5025 Transfer Stations. Before using the deep bath, please check
whether the silver/blue transfer arm is available. Older instruments are equipped with a silver/silver
transfer arm. The deep bath may not be used in combination with this transfer arm!
78
(→ Fig. 49-1)
of the transfer stations should be emptied after the
PAUSED
and
TS FULL
2
messages.
(→ Fig. 50-1)
TS FULL
only. The transfer arm