Display
Note
The following status messages are displayed only if the instrument is connected to a staining machine
(Leica ST5010 AutoStainerXL or Leica ST5020 Multistainer) via a transfer station.
Leica CV5030
Explanation
The Pick & Place module
loses the coverslip on the
way to the coverslipping
position.
1. The suction cups, the
cover slip sensor and/or
the skids on the Pick &
Place module are stuck
in place with coverslip
mountant.
2. The suction cups are
deformed.
3. The cover slip magazine
is inserted incorrectly.
4. The suction cups are
not correctly seated
on the Pick & Place
module.
Indicates that a specimen
slide holder that has been
transferred from the staining
machine is in the process of
coverslipping.
Operation
Action
As soon as the display reads
DROPPED, the instrument goes
automatically into
PAUSE
Notice: Move the dispenser into
the park position, move the Pick &
Place module towards the user and
remove the specimen slide. You
can now coverslip the specimen
slide manually. As an alternative,
immerse the specimen slide in a
compatible solvent and remove
the coverslip mountant, then re-
insert the specimen slide into the
instrument for coverslipping.
1. Clean the suction cups,
cover`slip sensor and/or the
skids of the Pick & Place
module using a lint-free cloth
moistened with compatible
solvent or replace them with
new ones
(→ p. 85 – 7.5.9
Cleaning and replacing the
suction
cups).
2. Replace the suction cups with
new ones
(→ p. 85 – 7.5.9
Cleaning and replacing the
suction
cups).
3. Check that the cover slip
magazine is seated correctly.
4. Check that the suction
cups are seated correctly; if
necessary, pull them off and
reattach.
Wait for the specimen slide to finish
processing.
5
CS
mode.
57