Checking for fredom of movement
1. Finally, ensure that the gripper
(→ Fig. 24)
2. If the edge of the specimen slide touches the lid of the brush holder, you can carry out a fine
adjustment after unscrewing the screws
so, use the available play in the bores.
Note
• Fill the container with 5 ml of solvent using the provided plastic pipette (→ Fig. 19-5). Regularly
check the level of the container. The solvent being used must be compatible with the coverslip
mountant.
Fig. 24
4.9

Connecting the power supply

L The electrical connection is on the left rear side of the instrument (→ Fig. 25-1).
L The unit is dimensioned for: 100 to 240 V AC voltage at 50/60 Hz.
Warning
The instrument MUST be connected to a grounded power socket.
• Only use a suitable cable that is appropriate for the local power supply (plug must fit on-site wall
outlet).
Connecting to power supply
1. Make sure that the coverslipping machine is switched off: The power switch
the front side must be in the "0"=
2. Insert a suitable power cable into the power input socket (→ Fig. 25-1).
Leica CV5030
(→ Fig. 24-1)
during its movement or that of the transported specimen slide (→ Fig. 24-2).
1
2
OFF
does not collide with the dispenser needle cleaner
(→ Fig. 21-3)
on the housing wall (→ Fig. 21). To do
position.
Instrument Setup
(→ Fig. 25-3)
on
4