4.11 Refilling consumables
• Fill the loading bath
should reach up to the specimen slide's label field.
• Fill the glass vial
mountant (approx. 10 ml). The dispenser needle must always be immersed in the solvent.
• The standard scope of delivery includes two different cover slip magazines with different widths
(22/24 mm). Fill and insert the magazine
used (possible cover slip lengths 40 - 60 mm)
38
1
2
1
Fig. 30
(→ Fig. 29-4)
with a solvent compatible with the coverslip mountant. The level
(→ Fig. 30-2)
in the rest position with a solvent compatible with the coverslip
Loading bath
• Open the loading door
out the loading drawer (→ Fig. 29-2).
• Insert the loading bath
drawer and insert the provided bath insert
(→ Fig. 29-3).
• The standard scope of delivery includes the
bath insert for the Leica 30 specimen slide
holder. The respective associated insert must
be used when using other specimen slide
holders (Leica metal 20 specimen slide holder,
Leica Sakura-type plastic 20 specimen slide
holder or specimen slide holders from other
manufacturers
Accessories).
2
(→ Fig. 30-1)
corresponding to the cover slip width being
(→ p. 99 – 9. Optional
(→ Fig. 29-1)
and pull
(→ Fig. 29-4)
into the
(→ p. 99 – 9. Optional
4
3
Accessories).