Advertisement

5
Operation
2
3
4
5
Fig. 45
1
Label field
2
LEN
= STROKE LENGTH CORRECTION
(length correction for the coverslip mountant application)
3
STK
= MOUNTANT STROKE
(length of the coverslip mountant application)
4
STP
= STROKE POSITION CORRECTION
(starting point of the coverslip mountant application)
5
CSP
= COVERSLIP POSITION CORRECTION
(cover slip placement position)
STP
and
CSP
must be matched with each other, since otherwise there is the danger that the coverslip
mountant could drip over the specimen slide edge into the instrument at the stroke start point, or
become stuck to the suction cup due to poor coverslip placement. In both cases, continuing smooth
instrument operation cannot be ensured.
Notes on fine-tuning the parameters STP, CSP,
• In most cases, the parameter
low-viscosity coverslip mountants (information is based on experimental values and may vary).
For lower values (negative value up to 0/+10), there is the danger that the coverslip mountant will flow
out over the specimen slide or coverslip edge.
On the other hand, if the starting point of the coverslip mountant stroke is still positioned too
far towards the specimen slide center, air pockets could form in the coverslip edge area after
coverslipping and drying of the coverslip mountant.
• The coverslip placement
relation to the specimen slide edge, which faces the label field). The settings are to be customized
72
1
LEN
STP
must be set in the positive range at approx. +20 to +40 for extremely
CSP
should be positioned before starting the coverslip mountant stroke (in
Version 3.4, Revision M