5.11 MENU B — parameter settings
Coverslip Position Corr — correction of the cover slip placement position
1
Fig. 43
Stroke Position Corr — start position correction of the coverslip mountant stroke
1
Fig. 44
Leica CV5030
The cover slip placement position can be modified using the CSP
(→ Fig. 43-1)
(cover slip placement position) parameter (→ Fig. 43).
• If the cover slip placement position is to be moved towards the edge of the
specimen slide, then select a negative value between –5 and –30.
• If the cover slip placement position is to be moved towards the label field, a
positive value between +5 and +30 has to be selected.
• The changes can be made in increments of 5.
One increment corresponds to 0.5 mm.
• The factory preset value is 0
The coverslip mountant stroke position can be changed using the
(→ Fig. 44-1)
(starting point for the coverslip mountant stroke) parameter.
• If the starting point for the coverslip mountant stroke is to be moved
towards the edge of the specimen slide, then select a negative value
between –10 and –100.
• If the starting point for the coverslip mountant stroke is to be moved
towards the label field, a positive value between +10 and +100 has to be
selected.
• The changes can be made in increments of 10.
One increment corresponds to 1.0 mm.
• The factory preset value is 0.
5
Operation
STP
63