Warning
• To avoid damaging the specimens specimen slides should never be inserted into a specimen holder on
which a clip has already been attached.
• The clip should only be attached after the specimen holder has been removed from the
HistoCore SPECTRA ST. Using the Clip during the staining process can cause collisions within the device,
which can lead to interruptions of the staining process.
• The specimen slide holder handle
before inserting it into the Leica CV5030.
Preparing the Leica CV5030 for the HistoCore SPECTRA ST Bath Insert:
1. Open the load drawer of the Leica CV5030
2. Remove the existing bath insert, check the bath for glass debris and clean if necessary.
3. Insert the HistoCore SPECTRA ST Bath Insert
Fig. 33
Attaching the Clip to the specimen slide holder:
1. Remove the specimen slide holder from the HistoCore SPECTRA ST.
2. Unfold the specimen slide holder handle and hold the specimen slide holder at an angle
(→ Fig. 34).
Fig. 34
Leica CV5030
(→ Fig. 35-1)
1
has to be removed from the specimen slide holder
(→ Fig. 32-4)
and pull the loading bath forward.
(→ Fig. 33-1)
into the loading bath.
5
Operation
45