Interrupting The Coverslipping Operation - Leica CV5030 Instructions For Use Manual

Robotic cover slipper
Hide thumbs Also See for CV5030:
Table of Contents

Advertisement

5
Operation
5.6

Interrupting the coverslipping operation

48
Note
If the space in the output magazine has already
been assigned completely but there are still
SPECIMEN SLIDES TO BE COVERSLIPPED
present in the loading bath, this is indicated by the
message
PAUSED
a signal tone.
The
CHECK LIFT
assigned output magazine has to be removed.
Insert the empty output magazines and only then
confirm the signal tone with the RESPOND button.
The
CHECK LIFT
disappear. The empty output magazines are
moved into the position required for retrieving
the coverslipping operation. The
remains on the screen.
The coverslipping operation can be continued by
pressing the START button.
Note
General note:
Full output magazines are automatically moved to
the output position with the following exceptions:
a) If just one 20-pack or 30-pack output magazine is
inserted
b) If just two 20-pack output magazines are inserted
c) If only part of the output magazine is occupied
after ending the coverslipping operation.
The LIFT button must be pressed in these cases.
Press PAUSE to interrupt the coverslipping operation.
• The currently processed specimen slide is completely
coverslipped and pushed into the output magazine,
accompanied by the message PAUSING.
• If this step is ended, the message
display.
A signal tone is emitted at the same time.
• Confirm the signal tone by pressing the RESPOND button.
and
CHECK LIFT
accompanied by
message is a note that the
message and signal tone
PAUSE
message
PAUSED
is shown in the
Version 3.4, Revision M

Advertisement

Table of Contents
loading

Table of Contents